H601-OLYMPUS-IX-SUSP-GLASS/FRAME

H601-OLYMPUS-IX-SUSP-GLASS/FRAME is a dual channel plate featuring Indium Tin Oxide (ITO) coated glass cased in a heated aluminium frame.

It provides a warm and optically transparent surface perfectly flush with the microscope stage with the best possible temperature uniformity and accuracy, achieved by controlling the temperature of both the glass and the metal frame. 

H601-OLYMPUS-IX-SUSP-GLASS/FRAME features anti-scratch, anti-fingerprint, chemically strengthened Dragontrail glass cased in rigid aluminum frame to ensures the highest resistance against breakage and scratches. 

Protected by 10 year warranty on glass damage.

It requires H401-T-CONTROLLER and uses both control channels (one for the glass and one for the flame).

Features

  • Flush with Olympus stages: IX-SUSP, GX-SVR, GX-SFR, BX3-SSU, IX2-SFR, IX2-SVL-2
  • Dual channel: heated Glass and heated Metal Frame
  • Temperature range: from ambient to 60°C
  • Accuracy on specimen: ±0.3°C
  • ITO glass coating ensures uniform glass heating
  • Glass thickness: 0.5 mm
  • Aluminum frame improves rigidity
  • Anti-scratch, anti-fingerprint, chemically strengthened Dragontrail glass
  • 10 year warranty on glass damage
H601-NIKON-TI-SR-GLASS/FRAME
H601-NIKON-TI-SR-GLASS
H601-OLYMPUS-IX-SUSP-GLASS-FRAME
H601-OLYMPUS-IX-SUSP-GLASS/FRAME
 
H601-OLYMPUS-IX-SUSP-GLASS-FRAME
Dimensions

 

H601-OLYMPUS-IX-SUSP-GLASSFRAME-IR
H601-OLYMPUS-IX-SUSP-GLASS/FRAME-IR

Temperature Uniformity

assessed via Infra Red Imaging

 

H601-GLASS:FRAME-STRUCTURE_420x280.jpg
DUAL HEATING TECHNOLOGY

Heated Glass - Heated Frame

Superior Temperature uniformity and stability obtained by heating both the glass and the metal frame

H401-T-CONTROLLER

2-Channel Temperature Controller with 4.3'' touch screen interface compatible with all H401 heating devices. One controller can control independently up to two items.

External temperature sensor (included) and self calibration routines allows easy calibration of the heating devices for the highest accuracy on sample temperature.

On board data logging and download procedures to USB drives allow to record temperature as a function of time and to log calibration events and alarms.

The controller is equipped with password protected routines to re-calibrate the sensors against certified thermometers. 

Features

  • T - range: from ambient to 60°C
  • T - accuracy on specimen: ± 0.3°C
  • 4.3'' touch screen interface
  • External T - sensor (included)
  • On-board Data logging and download routines to USB drive
  • On-board logging of calibration and alarm events
  • Alarm buzzer and External Alarm Connector
  • One controller can control independently up to two devices
  • Self calibration routines allow to improve temperature accuracy
  • Password protected procedure for periodic calibration against certified thermometer 
 
H401-T-CONTROLLER+Round_Glass_Plate_420x280.JPG
Touch Pad Interface 

Touch Pad Interface

Beautifully designed, simple to use and very compact.

T - range: from ambient to 60°C

4.3'' touch screen interface

Beautifully designed, simple to use and very compact.

External T - sensor (included)

On-board Data logging and download routines to USB drive

One controllers can control independently up to two devices

Self calibration routines allow to improve temperature accuracy

H401-T-CONTROLER+STAGE_TABLE_420x280.JPG
Clamp on stage T-Pad + Glass plate 

Take advantage of the dual channel option

One channel controls the metal T-Pad.

The other channel controls the Glass Plate.

H401-T-CONTROLLER+USB_420x280.JPG
On board Data Logging / Download to USB 

Data Logging

On-Board Data Logging - stored in local memory

A simple procedure allows to download data to USB.

 

Customer care is our top priority

CONTACT US

Okolab S.r.l.

FACTORY NAPOLI: Via A. Olivetti, 1 - 80078 Pozzuoli, NA, Italy

FACTORY ROVERETO: Via delle Zigherane - Edificio B08.01, 38068 Rovereto, TN, Italy

 

  • Luca Lanzaro Ph.D
  •   Phone: +39 081 8062624
  •   E-mail: info[at]oko-lab[dot]com

Okolab USA Inc.

417 Thorn St., Suite 309, Sewickley - PA 15143, USA

 

  • Lara Petrak
  • Mobile: +1 (650) 410 0756
  • Rob Davison
  • Phone: +1 (650) 228 6154
  •   E-mail: usa[at]oko-lab[dot]com

Okolab Shanghai Co., Ltd.

Room 2604, Jiande Apartment, No. 9 Jiande Road, Huangpu District, Shanghai 200025, China

 

  • Pony Tong
  • Mobile: +86 18500508458
  •   E-mail: tong[at]oko-lab[dot]cn
 

COPYRIGHT © OKOLAB S.R.L. ALL RIGHT RESERVED. CERTIFIED ISO 9001.